The HD-A513-1 ring press is one of the most commonly used products in the packaging industry. It serves Guangdong, Guangxi, Hainan and other regions and is also Haida's most marketed product. It has found that many customers have purchased the ring press and operated it. Not very familiar with the following Haida technicians to introduce you to the operation method of the lower ring press:

One. Prepare the sample:

1, the determination of the sample position: Select from the same batch of material scattered position without scratch samples.

2. Sample preparation: cut sample width 12.7±0.1 mm, length 152.0±0.2 mm

3, the number of samples: take more than 10 vertical and horizontal.

4, the standard state of the sample: before the test, the temperature was kept at 23±2°C, and the relative humidity was 60±5% for more than 24 hours.

note

n

G

Use GT-HY ring pressure tester to cut the standard sample (optional)

n No sharp edges or other defects affecting the measurement result are allowed on the edge of the section.

n Glove-free hands are required to touch the test specimen.

n Samples with long sides parallel to the longitudinal direction are used to determine the lateral crushing strength.

n Samples with long sides parallel to the longitudinal direction are used to determine the lateral crushing strength.

Two. Upper and lower safety protection device settings:

1, the role of upper and lower limit devices, mainly include:

A Prevent the upper platen from dropping to impact the lower platen or test flat jig and damage the load cell

B As long as the pressure plate on the touch plate, touch the upper limit button or lower limit setting button, the upper pressure plate will automatically stop.

C can adjust the upper and lower pressure plate or the pressure plate and test fixture spacing.

2, first fixed upper and lower limit fixed block in place.

3. The upper limit setting button is set to a position sufficient for the upper pressure plate to place the sample test jig or sample, and the lower limit setting button should be set on the upper pressure plate to descend to the upper or lower pressure plate or the upper pressure plate. Test fixtures do not impact each other.

4. To loosen the screw of the upper or lower setting button, you can move the setting button. To move to positioning, you must tighten on the touch bar.

caveat

n

!

The upper and lower limit fixed blocks have been adjusted before delivery, not necessary, do not move.

n After replacing different test fixtures, the position of the upper and lower limit setting buttons must be reset.

n When the lower limit setting button is locked, it should be locked tightly to prevent it from crashing and crashing into the heavy load.

n After setting, you should test whether the protection function of the upper and lower limit setting buttons is normal.

Third, the operation steps

1. According to the thickness of the sample, choose the diameter of the inner plate to be placed on the sample holder (indicated on the inner plate for the sample thickness in mm), carefully insert the sample into the sample holder, and make sure to insert the sample into the bottom

3. The machine platform must be connected to a 220V ± 5% 50HZ power supply and good grounding protection measures must be provided.

4. After turning on the power switch on the panel, the panel is powered up and the monitor will automatically open the program.

5. Check whether the stroke protection device of the machine is valid and the position is reasonable.

6. Press the zero key to set it to zero, press the “peak” key, press the “down” key, and after the surface of the desired test article, press the “test” key to test.

7. Place the test specimen in the middle of the platen, press the zero and peak keys, and continue to press the down key to start the test.

8. Do not allow direct contact with the upper and lower platens or test with metal-type test pieces.

9. In the test process, if there is an abnormal situation, press the stop button and immediately notify the maintenance personnel to test. Can be used only after troubleshooting.

10. The display accepts the pressure sensor signal from the top of the upper platen and displays the current pressure value. Press the zero key to clear

The number on the display. Press the peak key to record the maximum value of the test during the test.

11. Date, time, breakpoint ratio and setting parameter settings:

step

operating

display

Instructions

1 .

After power on

Press [Function]

【d 060311】

After power on, press and hold [Function]

2.

Press [Function]

Press 【Print】

Press [peak]

【d 060322】

Date setting

If you need to modify, press [Function] to shift, press [Print] to increase the number. Press

[Peak] key confirmation

3.

Press [Function]

Press 【Print】

Press [peak]

【t 101208】

time setting

If you need to modify, press [Function] to shift, press [Print] to increase the number. Press

[Peak] key confirmation

4.

Press [Function]

Press 【Print】

Press [peak]

[b 85]

Breakpoint ratio setting

If you need to modify, press [Function] to shift, press [Print] to increase the number. Press

[Peak] key confirmation

5.

Press [Function]

Press 【Print】

Press [peak]

[A 50.00]

Quantitative value setting

If you need to modify, press [Function] to shift, press [Print] to increase the number. Press

[Peak] key confirmation